Design of Fully Integrated Wireless CMOS MEMS Device for Intraocular Pressure Measurement
No Thumbnail Available
Files
Date
2008-05-10
Authors
Journal Title
Series/Report No.
Journal ISSN
Volume Title
Publisher
Abstract
The thesis research presented here explores the design and fabrication of wireless Integrated CMOS MEMS device for continuous intraocular pressure measurement. A complete study of the existing solutions for sensing intraocular pressure is made and the problems involved in them are identified in the report. A simple, low power, less invasive solution for the application is then proposed. Electromagnetic modeling of the inductive link is done and optimum dimensions for the antenna are finally suggested. Complete design of all the components of the system such as control circuit, power conversion and conditioning circuits, reference circuits necessary for the reliable functioning of the device is presented. Integration of MEMS sensor along with the analog circuitry and antenna is detailed. Techniques for packaging the device in order to meet the standards for safe operation are then presented. The report has dedicated chapters on review of the existing sensors, Inductive link design, CMOS control and Power conditioning circuit, Sensor Integration and Packaging. The principles, methods, systems and circuits described in this thesis are applicable to other implantable wireless systems which have stringent requirements on power and space.
Description
Keywords
Intraocular pressure measurement, CMOS mems
Citation
Degree
MS
Discipline
Electrical Engineering