Wells, Heather. Masters of Environmental Assessment. Assessing Low-Cost Purple Air
Particulate Matter Sensors
With technological advances, low-cost light scattering particulate matter (PM) sensors have
come into more prevalent use. They operate by sensing laser light scattering by particles. The
scattered light correlates to particle count, and then sensor algorithms can determine the size-
selective PM concentration based on the intensity of the light and assumptions about average
particle density. Low-cost sensors are not as accurate as Federal Reference Methods (FRMs),
with concentrations having lower accuracy and precision. In order to be able to evaluate air
quality using low-cost sensors, it is necessary to test reproducibility of the data. In this study,
four PurpleAir low-cost sensors were collocated with a Federal Reference Method to measure
PM in a chamber with a wood stove as the source. A data correction scheme for the PurpleAir
sensors was developed. While the PurpleAir sensors had high correlations with the tapered
element oscillating microbalance (TEOM) reference instrument, their concentrations were
roughly 50% lower than those produced by the TEOM. A limitation of the study was that the
sensors were only tested in a chamber environment with no varying environmental conditions.
The chamber experiment should be repeated seasonally to account for any sensor drift or
changes from environmental factors and for a number of sources and ambient conditions.