A Wideband Power Efficient Inductive Link for Implantable Biomedical Devices using Multiple Carrier Frequencies

dc.contributor.advisorMaysam Ghovanloo, Committee Chairen_US
dc.contributor.authorAtluri, Sureshen_US
dc.date.accessioned2010-04-02T18:05:53Z
dc.date.available2010-04-02T18:05:53Z
dc.date.issued2006-11-12en_US
dc.degree.disciplineElectrical Engineeringen_US
dc.degree.levelthesisen_US
dc.degree.nameMSen_US
dc.descriptionNorth Carolina State University Theses Electrical and Computer Engineering.
dc.description.abstractThis thesis presents a novel inductive link for wireless transmission of power and data to biomedical implantable microelectronic devices using multiple carrier frequencies. Achieving higher data bandwidth without compromising the power efficiency is the driving force to use two separate carriers. Two separate pairs of coils have been utilized for inductive power and forward data transmission. One major challenge, however, is to minimize the interference among these carriers especially on the implantable side, where size and power are highly limited. Planar power coils with spiral shape are optimized in geometry to provide maximum coupling coefficient, k. The data coils are designed rectangular in shape and wound across the power coils diameter to be oriented perpendicular to the power coil planes. The goal is to maximize data coils direct coupling, while minimize their cross-coupling with the power coils. The effects of coils geometry, relative distance, and misalignments on the coupling coefficients have been modeled and experimentally measured. A prototype system is built to test the functionality of the approach. Frequency Shift Keying (FSK) is used to transmit data using two carriers at 6 and 12 MHz. Power is transmitted at 500 kHz. We achieved power efficiency of 43% at a nominal distance of 10 mm. We also achieved very minimal power interference on to data using the devised coil arrangement. Initial results show the functionality of the design. The inductive link needs more tuning to reduce the Bit Error Rate (BER) for successful operation of an implantable micro stimulating system. In addition a novel back telemetry link is developed to transmit data back from the implant. This data gives information related to received power, electrode impedances etc, to the external part of the system operation and helps achieve a more stable closed loop system. We have also designed and developed a Manchester encoded FSK Data demodulator circuit to be used in our multi carrier system but have not tested it.en_US
dc.formatThesis (M.S.)--North Carolina State University.
dc.identifier.otheretd-07062006-194716en_US
dc.identifier.urihttp://www.lib.ncsu.edu/resolver/1840.16/1656
dc.rightsI hereby certify that, if appropriate, I have obtained and attached hereto a written permission statement from the owner(s) of each third party copyrighted matter to be included in my thesis, dissertation, or project report, allowing distribution as specified below. I certify that the version I submitted is the same as that approved by my advisory committee. I hereby grant to NC State University or its agents the non-exclusive license to archive and make accessible, under the conditions specified below, my thesis, dissertation, or project report in whole or in part in all forms of media, now or hereafter known. I retain all other ownership rights to the copyright of the thesis, dissertation or project report. I also retain the right to use in future works (such as articles or books) all or part of this thesis, dissertation, or project report.en_US
dc.subjectImplantable micro devicesen_US
dc.subjectinductive linken_US
dc.subjectclass E power amplifieren_US
dc.subjectback telemetryen_US
dc.titleA Wideband Power Efficient Inductive Link for Implantable Biomedical Devices using Multiple Carrier Frequenciesen_US
dcterms.abstractKeywords: Implantable micro devices, inductive link, class E power amplifier, back telemetry.
dcterms.extentix, 84 pages : illustrations (some color)

Files

Original bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
etd.pdf
Size:
1.87 MB
Format:
Adobe Portable Document Format

Collections